Vial washing machine is the
one of critical Equipment in the aseptic manufacturing process which plays a crucial
role in preparing containers for aseptic filling. To meet regulatory
requirements, it is essential to conduct a thorough qualification of the vial
washing machine. This article aims to provide a comprehensive guide for the qualification
vial washing machine, covering all key concepts, processes, and regulatory
considerations.
Understanding Vial Washing
Machine Principle:
Vial Washing Machine are
two type liner and rotatory type. In both type the vials are cleaned in below
mentioned Steps:-
1st Step:
External/Internal Wash with recirculated water
2nd Step: External/Internal
Compressed Air Blow
3rd Step:
Internal Wash with Purified water
4th Step: External/Internal
Compressed Air Blow
5th Step External/Internal
wash with Water for injection
6th Step External/Internal
Compressed Air Blow
Performance Qualification
(PQ) and Requalification:
Performance of the vial
washing machine evaluated in simulated production environment.
Challenge Test Planned for Vial Washing Machine
1.0 Spiked
Vial with Riboflavin dye challenge for Proper cleaning and florescence observed
after cleaning under the UV cabinet. Use can use the concentration between 0.08g/L
to 0.22g/L
2.0 Spiked
Vials with Charcoal Slurry for determination of Visible and sub visible black
particle after washing of vial. Vials Challenged in the white background for Visible
Particles and for sub visible analysed through Liquid Born particle analyser.
3.0 Spiked Vials with Talcum powder for
determination of Visible and sub visible white particle after washing of vial.
Vials Challenged in the black background for Visible Particles and for sub
visible analysed through Liquid Born particle analyser.
4.0 Spiked
Vials with glass particles for determination of Visible glass particle after
washing of vial. Vials Challenged in the black and white background for Visible
glass Particles
5.0 Spiked
Vials with NaCl Solution for determination of chemical contamination after
washing of vial. Cleaned Vials are test in QC for any NaCl residue
contamination.
6.0 Residual
water content after blowing of compressed air inside and outside of vial to assure
the vial having less than 0.4g so that it will evaporate during dehydrogenation
tunnel
7.0 Endotoxin
Spiked vial challenge for 1 log reduction after washing.
Before Starting the challenge test 100 ml Sample for liquid born particle count analysis shall be collected for Recirculated Water, Purified Water, WFI purge with Compressed Air.